MDD01M-12N

Stock number: C009618

Manufacturer: MEANWELL


€4.41 incl. VAT €3.64 Excl. VAT


To order

To place an order, fill in the inquiry form and our sales team will contact you.

In our warehouse 0 pcs External warehouse 0 pcs

We will confirm the exact delivery date in our offer.


Product description

The MEAN WELL MDD01M-12N is a miniature isolated DC/DC converter from the updated MDD01-N series, designed for medical electronics, measurement instruments, analogue circuits, industrial control systems and galvanically isolated power applications. From a nominal 12 V DC input, it generates two symmetrical outputs of +12 V DC and −12 V DC with a common reference point. Each output rail provides up to 42 mA, with a combined rated output power of 1 W. Its compact SIP7 package, measuring 19.7 × 7.1 × 11.5 mm, and through-hole PCB mounting make it suitable for electronic equipment requiring space efficiency, symmetrical power rails and high isolation strength.

A major advantage of the MDD01M-12N is its reinforced input-to-output galvanic isolation, rated at 6,000 V DC. The converter is designed for applications with enhanced medical electrical safety requirements, supporting an isolation level of 1× MOPP / 2× MOOP according to IEC 60601-1, subject to defined system conditions. Its exceptionally low patient leakage current below 2 µA and typical 10 pF isolation capacitance are beneficial in sensitive measurement and medical equipment. Galvanic isolation allows power domains to be separated and helps limit unwanted potential differences between circuits. The symmetrical ±12 V supply is suitable for operational amplifiers, analogue filters, instrumentation amplifiers and signal conditioning circuits.

The MEAN WELL MDD01M-12N operates from a 10.8–13.2 V DC input, corresponding to ±10% of its nominal 12 V rating. It achieves a typical efficiency of 80% and operates at a switching frequency of approximately 30 kHz. Its outputs are unregulated, meaning that the actual output voltages depend on input voltage and the loading of each rail. The manufacturer specifies ±5% output voltage accuracy, ±10% load regulation and maximum ripple and noise of 100 mVp-p under defined test conditions. The recommended load range for each output is 5–42 mA, with a maximum capacitive load of 100 µF per output. A typical 18 mA no-load input current and 104 mA full-load input current facilitate accurate sizing of the primary power supply circuit.

The updated MDD01-N series supports an extended operating temperature range of −40 to +95°C, subject to derating, and uses fanless natural convection cooling. Integrated continuous short-circuit protection enables automatic recovery once the fault is removed. The manufacturer recommends a 500 mA slow-blow input fuse. The module is housed in a non-conductive plastic enclosure with a UL94 V-0 flammability rating, an approximate weight of 2.8 g and a standard SIP7 pin arrangement. The manufacturer specifies compliance with IEC/UL 62368-1 and IEC 60601-1 safety requirements and provides a 3-year warranty. Combining 1 W output power, symmetrical ±12 V outputs, 6 kV DC isolation, 80% efficiency and a compact SIP7 package, the MDD01M-12N is a suitable solution for professional medical, measurement and industrial electronic applications.

Key features

  • Product type: isolated PCB-mount DC/DC converter

  • Manufacturer series: MDD01-N

  • Construction: encapsulated SIP7 module

  • Number of outputs: 2, symmetrical

  • Nominal input voltage: 12 V DC

  • Input voltage range: 10.8–13.2 V DC

  • Input voltage tolerance: ±10%

  • Output voltage 1: +12 V DC

  • Output voltage 2: −12 V DC

  • Symmetrical output voltage: ±12 V DC

  • Rated output current 1: 42 mA

  • Rated output current 2: 42 mA

  • Load range per output: 5–42 mA

  • Combined rated output power: 1 W

  • Output regulation: unregulated converter

  • Typical efficiency: 80%

  • Switching frequency: approximately 30 kHz

  • Output voltage accuracy: ±5%

  • Ripple and noise: 100 mVp-p maximum

  • Line regulation: 1.2% per 1% input voltage variation

  • Load regulation: ±10%

  • Maximum capacitive load: 100 µF per output

  • Typical no-load input current: 18 mA

  • Typical full-load input current: 104 mA

  • Input filter: internal capacitor

  • Galvanic isolation: yes

  • Insulation type: reinforced

  • Input-to-output withstand voltage: 6,000 V DC

  • Isolation resistance: 100 MΩ at 500 V DC

  • Typical isolation capacitance: 10 pF

  • Isolation level: 1× MOPP / 2× MOOP under manufacturer-specified conditions

  • Patient leakage current: less than 2 µA

  • Short-circuit protection: continuous, automatic recovery

  • Recommended input fuse: 500 mA slow-blow

  • Cooling method: natural convection

  • Operating temperature: −40 to +95°C, subject to derating

  • Maximum case temperature: +105°C

  • Operating humidity: 20–90% RH, non-condensing

  • Storage temperature: −55 to +125°C

  • Temperature coefficient: 0.02%/°C from 0 to +90°C

  • Vibration resistance: 2G

  • Case material: non-conductive plastic

  • Case flammability rating: UL94 V-0

  • Mounting type: through-hole (THT)

  • Package: SIP7

  • Number of populated pins: 5

  • Dimensions: 19.7 × 7.1 × 11.5 mm

  • Approximate weight: 2.8 g

  • Safety standards: IEC/UL 62368-1, IEC 60601-1

  • Manufacturer's warranty: 3 years

Target applications

  • Medical electronics and diagnostic instruments

  • Galvanically isolated analogue power supplies

  • Operational amplifiers requiring symmetrical ±12 V rails

  • Instrumentation amplifiers and precision measurement systems

  • Analogue filters and signal conditioning circuits

  • Medical monitors and diagnostic equipment

  • Isolated communication and data interfaces

  • Industrial controllers and sensor modules

  • Telecommunications and data equipment

  • OEM manufacturing of medical and measurement systems


Parameters

Product parameters

Description: DC/DC converter unregulated
Case: SIP-7
Operating temperature: -40~95 °C
Dimensions: 19,7x7,1x11,5 mm
Output voltage DC: -12 V
Output current: ±42 mA
Series: MDD
DC input voltage: 10,8~13,2 V
Power of convertor: 1 W
Number of outputs: 2
Isolation: 6000 V
Weight: 2,8 g
Efficiency (typ.): 80 %
Model series: MDD01-N
Approval agency: CE, UKCA, cURus, CB, EAC
Protection: Short Circuit Protection (SCP)

Packaging and legislation

Typ balení: Tube
Původ: TW
Small packing quantity: 24
Large packing quantity: 2160
Customs nomenclature: 85044095